Views: 0 Author: Site Editor Publish Time: 2025-03-28 Origin: Site
1. Pre-Calibration Check
Inspect the instrument condition.
Clean the sample chamber and sensors.
Verify temperature and pressure stability.
2. Reference Standard Calibration
Use Certified Reference Materials (CRMs) or calibration gases with known calorific values.
Compare the analyzer’s reading with the certified value.
Adjust the analyzer settings if deviations are detected.
3. Validation with Known Samples
Run a known fuel sample (e.g., natural gas or coal) with a defined calorific value.
Verify the analyzer provides accurate and consistent results.
4. Documentation and Certification
Record the calibration data.
Generate a calibration certificate for compliance purposes.
Store calibration records for audit and traceability.
Even if the analyzer is on a regular calibration schedule, certain indicators suggest early calibration is required:
Inconsistent or drifting results: Deviations from expected values.
Frequent measurement errors: Fluctuations in readings.
Longer stabilization times: The device takes longer to display stable readings.
Poor repeatability: Inconsistent results from the same sample.
Environmental changes: Variations in temperature, humidity, or pressure.
Auto Lubricating Oil HTHS Dynamic Viscosity Tester Manufacturer
Apparent Viscosity Analyzer at High Temperature And High Shear Rate
High Temperature Foaming Characteristics of Lubricating Oils Tester
Foaming Characteristics of Lubricating Oils Tester Manufacturer
Lubricating Oils Foaming Characteristics Tester Manufacturer
Water Separability of Petroleum Oils And Synthetic Fluids Tester
Evaluating Wear Resistance with The Lubricant Anti‑Wear Performance Tester
Quantifying Volatility with The Lubricating Oil Evaporation Loss Tester
China Fretting Wear Protection by Lubricating Greases Tester
Characterizing Foam Behavior with the Lubricating Oils Foaming Characteristics Tester
Determining Flammability with The Automatic Open Cup Flash Point Tester